This feature is available in software version 0.10 or later.
Connection Overrides
In some cases a Test Object may not be natively compatible with an EST. This can happen if the not all sockets on the Test Object fit available sockets on the EST. To test this object you must create an adapter.
To use this adapter, with older generations EST, you would modify the Test Object definition as if it already included the adapter. However, this can become problematic with multiple dissimilar ESTs and may result in multiple Test Object Definitions per EST. This approach complicates serial number tracking.
A Connection Override allows you to manually override the Test Object's sockets to available sockets on a specific EST, making the object forcibly compatible with that EST.
At the assumption that you keep a suitable adapter complementary to the created override.
The Test Object can then be tested with the EST using the adapter, without modifying the original Test Object definition.
The following section describes how Test Object compatibility with registered EST devices is determined and displayed, and how to configure a Connection Override for Test Objects that are not natively compatible. It covers both the Office application and the Test application.
Safety
As this feature involves creating custom adapters these must be constructed or bought with safety in mind.
Custom adapters must only be made by qualified personnel. Each custom adapter must be labeled with a name, date, and serial number to enable proper documentation of periodic inspections.
- Do not create dangerous male-to-male or female-to-female adapters.
- Do not create adapters that allow for the connection of live voltage to the EST.
- Do not create adapters that allow for the connection of external test equipment to the EST.
- Do not create adapters with exposed conductive parts, only use existing CE-compliant connectors or plugs.
- Do not create adapters that reduce the voltage rating, eg: Schuko to Ethernet.
When, for ease of handling, adapters are created with a reduced wire gauge than what is required for the rated current of the used sockets, ensure the adapters are clearly labeled to indicate their limitations and purpose.
EMRI is not responsible for any hazards or damages resulting from the creation or use of custom adapters.
Office Application
1. Compatibility Overview
Every Test Object detail page shows a Compatibility card indicating how well the object matches the sockets of the customer's registered EST devices.
There are three possible states:
| State | Description |
|---|---|
| Compatible | The Test Object matches natively with all registered EST(s). No action required. |
| Partially Compatible | The Test Object is natively incompatible with one or more registered EST(s), or has been made compatible for some of them through a Connection Override. |
| Incompatible | The Test Object cannot currently be used with the registered EST(s). |
The card also distinguishes between:
- Native — compatibility based on the object's own socket definitions, without any manual intervention.
- Forced — compatibility achieved through a manually configured Connection Override.
The "Forced" distribution box is natively incompatible, but has been made forcibly compatible using a Connection Override. The compatibility card shows it is compatible with 1 of the 5 registered EST(s).
2. Viewing Compatibility Details per EST
To see exactly which registered EST(s) a Test Object is compatible with — and why it may be incompatible with others — open the EST Compatibility Details pop-up via the expand arrow in the top-right corner of the Compatibility card.
The pop-up lists all registered EST(s) on the left, with a green check or red cross indicating compatibility. Selecting an EST on the left shows its details on the right, including:
- Whether the match is Native or Forced
- Which specific sockets on the Test Object are incompatible with that EST
Selecting "EST2300P001" shows it is incompatible, listing the two sockets ("CEE 16A 24V 2p Connector" and "XLR 1A 50V 3p Outlet") that could not be matched.
3. Connection Overrides
When a Test Object is not natively compatible with a given EST, a Connection Override can be configured to manually map the Test Object's sockets to the EST's available sockets. This makes the object forcibly compatible with that specific EST.
3.1 Viewing existing overrides
The Connection Override action on a Test Object's detail page opens a list of all overrides currently configured for that object, filterable by EST serial number.
Overrides configured for the "Forced" distribution box, filtered to EST serial number EST2604A002. One override, "XLR - Shuko", is listed.
From this list, an existing override can be edited or deleted using the Edit / Delete buttons.
3.2 Creating or editing an override
Selecting Add (or Edit on an existing entry) opens the override editor. To configure a new override:
- Select the target EST serial number from the dropdown in the top-right.
- Enter a name for the override.
- Map the Test Object's sockets (shown on the left, following the object tree structure) to the corresponding EST sockets (listed on the right, filterable by name).
- Click Save mapping.
Sockets marked required must be mapped for the override to make the object compatible with the selected EST; mapping the remaining, non-required sockets is optional.
Mapping the "XLR 1A 50V 3p Outlet" socket (marked "required") to an available EST socket. Available EST sockets can be filtered using the search box on the right, and are dragged onto the "Drop EST part here" target.
4. Fully Compatible Test Objects
Test Objects that are natively compatible with all registered ESTs do not require any override configuration. The Compatibility card shows a green check and "Compatible with all (n) registered ESTs".
Socket matching for these objects is performed automatically; a Connection Override can still be added manually if a specific, non-default mapping is desired.
The "Medium Dist. Box" is natively compatible with all 5 registered ESTs — no override is required.
Test Application
5. Selecting a Test Object and Test Plan
On the Test selector page, a Test Object instance (identified by its serial number, via search or barcode Scan) is selected from the list on the left, and a test plan is chosen from the Available test plans panel on the right.
The "Medium Dist. Box" instance (serial 19238436) is selected, along with the "NEN 3130 Cert." test plan. Clicking Start test proceeds to the socket connection step.
6. Socket Connection
After selecting an object and test plan and clicking Start test, a Socket connection pop-up is shown, offering up to three options:
| Option | Description |
|---|---|
| Auto Connect | Automatically creates socket connections based on native compatibility. Disabled if the selected object is not natively compatible with this EST. |
| Connection Override | Choose an existing connection override profile from the list configured for this object/EST combination. |
| New Override | Create a new connection override profile for this object, targeting this EST. Only shown to users with System Engineer permissions or higher. |
If no connection overrides exist for the Object/EST combination, the app skips this pop-up and proceeds straight to the test phase using Auto Connect.
The "Medium Dist. Box" is natively compatible with this EST, but the pop-up still appears because an override profile has also been defined for this object/EST combination — giving the user a choice.
6.1 Choosing an existing override
Selecting Connection Overrides opens a list of the override profiles available for this object and EST. Selecting Use on a profile continues to the test phase using that mapping.
One override profile, "Inlet Shuko -> Universal", is available and can be applied via Use.
6.2 Objects that are not natively compatible
For an object that is not natively compatible with this EST (e.g. "Forced"), Auto Connect is disabled — an override profile must be selected instead.
Auto Connect is greyed out with "Auto Connect is not available for this test object and EST combination." The user must use Connection Overrides or New Override instead.
6.3 No override exists yet
If the selected object is incompatible with this EST and no override has been configured yet, a warning is shown. Users with System Engineer permissions or higher are offered a Create connection override shortcut; other users only see the warning.
"This object has sockets that do not connect to this EST." Start test is disabled until a connection override is created.
6.4 Creating a new override from the Test application
Selecting New Override (from the Socket connection pop-up, or via Create connection override on the warning) opens the override creation screen as a modal. It works the same way as the Office application's override editor, except the target EST is pre-selected to the current device and cannot be changed.
A nickname must be entered before the object tree becomes available for mapping. The EST is fixed to "EST2604A002" — the device the Test application is running on.
Creating Adapters
When creating adapters for testing, you must ensure this adapter is safe. It should not pose any risk to the equipment, the user, or the testing environment.
If for ease of handling you decide to create this adapter with a non-conform wire guage for the used socket current rating, you must label it clearly to indicate testing use only.
Pin Mapping
On an override socket the software connects the pins of the original socket as best as possible.
Any Phase L1, L2 or L3 pin will be mapped to the corresponding phase pin on the override socket. Any Neutral (N) pin will be mapped to the Neutral pin on the override socket. Any Protective Earth (PE) pin will be mapped to the Protective Earth pin on the override socket. Any other pins will be mapped according to their pin numbers.
This allows overrides to increase the amount of pins, skipping pins that are not present on the Test Object.
Examples:
| Test Object [CEE 32A 4p] | Override Socket [CEE 32A 5p] |
|---|---|
| L1 | L1 |
| L2 | L2 |
| L3 | L3 |
| N (not used) | |
| PE | PE |
| Test Object [XLR 3p] | Override Socket [XLR 5p] |
|---|---|
| Pin 1 | Pin 1 |
| Pin 2 | Pin 2 |
| Pin 3 | Pin 3 |
| Pin 4 (not used) | |
| Pin 5 (not used) |