Maintenance
To keep the EST in perfect safe working order it is essential to perform regular inspections and calibrations. When the maintenance schedule is followed the risk of downtime is minimized.
| Maintenance | Interval |
|---|---|
| Self Diagnostic | Every 1 month |
| Inspection | Every 6 months |
| Calibration | Every 1 year |
| Refurbish | Every 8 years |
Self Diagnostic
The EST internal self-diagnostic quickly verifies many internal components and their integrity.
Any SystemConfigurator can perform this test.
Run at least the Internal Resistance Check after changing instruments.
Inspection
During an inspection, the entire EST is checked for damaged components and wiring. This inspection may be performed by the person responsible for the machine.
- Visually inspect the cabinet and footing.
- Visually inspect all test sockets for dirt, damage and wear.
- Visually inspect all test cables for damage.
- Test the emergency stop button.
Contact EMRI for replacement parts or service.
Avoid excess use of cleaning solvent sprays to enhance socket contact quality. Replace worn out sockets instead.
Calibration
It is possible to ship the measurement instruments to a calibration facility. This usually takes ~2 weeks.
It is recommdend to perform the inspection and self-diagnostic routine at the same time.
Run a Measurebus Selftest after changing instruments.
Swap service
To limit downtime, it is recommended to use our swap service. We will replace the instruments with temporary units while yours are sent to a calibration facility.
When the instruments are calibrated they are swapped out again.
~1 day of downtime.
Contact EMRI to request this service.
Refurbish
After many years of reliable service a refurbishment is required.
During refurbishment the worn relays are replaced, instruments replaced when neccesary and safety system are validated.
Contact EMRI to request this service.