Transport and Installation
The following instructions cover the transport, installation and placement guidelines of the EST.
If a printout of this page is sent in advance, the full user manual can be found here.
Arrival and Initial Handling
An EST arrives in one or more shipments on a custom pallet wrapped in plastic.
These shipments should be moved indoors immediately to prevent damage from weather.
- Please move the EST indoors immediately.
- Due to the width and uneven load distribution of the EST ensure forklifts widen the forks as much as possible.
- For internal (indoors) transport take the EST off the pallet and place on it's transport dollies.
- Do not transport the EST on your own, always have at least one other person assisting.
- The EST may not fit through standard 90cm wide doorways.
- On the final location, ensure sufficient space in front and behind for assembly and inspection. See Placement.
Some shipments of the EST are top-heavy and may tip over if not handled carefully.
Internal Transport
Refer to the Technical Specifications for the dimensions and weight of the EST.
Care must be taken when transporting an EST.
- Keep dry
- Do not drop
- Keep upright
- Separate cable units may be top-heavy
A full three-module EST (cable unit, cabinet, and table) can be shipped without difficulty.
Larger configurations will be shipped disassembled. See Assembly. If you are shipping an EST, we recommend splitting it between two cable units. See Disassembly.
The dimensions of a fully assembled EST are too wide to load on tail lifts. Side loading is required for transport.
Please include possible pallet height when planning transportation.
Forklift only
Do not hoist
The frame is not designed to be lifted. It may bend or get damaged if hoisted.

The EST may be carefully moved using a forklift.
Ensure that the forklift forks are spread as far apart as possible for maximum stability.
Use the forklift to remove the transport pallet and place the EST on its transport dollies.
Avoid using the transport dollies outdoors.
Transport dollies
An assembled EST can be moved by placing it on transport dollies.
Every shipment includes sufficient transport dollies. Keep these for future use.
Ensure that at least two people are present.
- Drive a pallet jack near the center or side module.
a. Ensure that the pallet jack does not damage anything behind the EST. - Lift one side of the EST.
- Place a dolly under the EST. (⚠️ Ensure that the dolly does not bind on the mounting plates.)
- Carefully lower the EST.
- Repeat for the other side. Very large ESTs may require more than two dollies.
Ensure that the mounting feet do not interfere with the wheels or get lost during transport.
If necessary, temporarily remove the bottom parts of the feet; they snap off and on easily.
Follow the steps above in reverse to remove the transport dollies and move the EST to its final position.
Location
Once the EST has been transported to its final location, assemble and position it correctly.
The EST must only be used indoors. The recommended temperature is 15 – 40 °C.
Keep it plugged in to prevent condensation. The internal heaters work at around 15 °C.
The power inlet (CEE 16A 5-pin) is on the lowest shelf on the table side.
If your EST arrived in separated modules, do not plug in the mains power until assembly is complete.
After transport, let the EST warm up for at least 30 minutes before switching it on.
Safety Zone
The EST automatically applies power to equipment being tested. Create a safety zone around it to protect anyone not using the EST.
Keep the area 3 m in front of the EST and 1.5 m on each side clear.
Mark this area with floor-marking tape or similar yellow lines.

Test objects connected to the EST should never be connected to an external power source.
We strongly recommend removing any free CEE or equivalent power connectors from the exclusive zone.
EST configurations may differ, so adapt these keep-out and safety zones to your situation.
If in doubt, contact EMRI for advice.
Maintenance Access
For maintenance and service, access to the cabinets of the EST is required.

- Zone A: The rear cabinets should be readily accessible during service.
There are no customer-serviceable components in the rear cabinets. - Zone B: Keep this area clear so the main cabinet door can be opened during general operation and service.
The EST may be placed against a wall.
Keep equipment such as pallet jacks nearby so you can pull the EST forward to access the rear cabinets for service.
Assembly
Some components of the EST are detached during shipping and must be assembled on site.
Larger EST configurations may arrive partially assembled and require on-site assembly and wiring.
Ensure that the EST is unplugged when assembling or disassembling modules.
Display
During shipping, the display is detached. To attach it to the VESA mount, first remove the locking screw and detach the bracket from the arm. The screw is circled in red in the photo below. Then attach the VESA adapter to the display using the included M4x15 screws.


Reattach the display to the arm, secure it with the locking screws, and connect the cables. Adjust the display mount and secure it with its wing nuts. You can also adjust the display height by moving it along the slot in the extruded profile.
When adjusting the touchscreen position, ensure sufficient cable slack is available. Overextending the screen may strain or disconnect the cables, potentially causing damage or loss of functionality.
Joining Modules
The modules of an EST are joined by four coupling plates: two at the top and two at the bottom.
When delivered, each module has half of the coupling plates on each side; the fasteners are stored in the nearby frame.
Follow these steps to safely join the modules.
The plates are set back from the edge so you can see whether the hammer nuts are correctly rotated.
Please note that individual modules have feet on only one side and may tip over when detached.
| Step | Action | Illustration |
|---|---|---|
| 1 | Ensure that no fasteners are inserted in the top coupling plate. | ![]() |
| 2 | Ensure that the bottom fasteners are aligned to receive the module. | ![]() |
| 3 | Lift the side of the module without bottom coupling plates. | |
| 4 | Move the other module snugly against the stationary module, aligning the front edges. | |
| 5 | Slowly lower the lifted side onto the bottom coupling plates. The aligned fasteners should slide into the slots in the frame. | |
| 6 | At the top, insert the hammer nuts into the coupling plates. | |
| 7 | Tighten all fasteners to secure the modules together. Ensure that the hammer nuts are rotated correctly and engage with the slots in the extrusion profile. |
Feet Adjustment
When the EST is in its final position, level the feet to ensure stability.
The cabinet is equipped with adjustable feet to keep it level and stable.
Ensure that all feet touch the floor and that the frame is straight. Do not overtighten the feet, as this may damage the threads.

Do not screw the levelling feet too far inward.
Doing so may damage components or surfaces above the adjustment screws.
Wiring
Modules are wired with multiple cables running through a cable hose. Each cable connects to its corresponding terminal inside the cabinet; the cable markings and conductor numbers identify the correct terminals.
Incorrect wiring may damage equipment or endanger people.
Contact EMRI if you are unsure or if wiring inside a locked cabinet is required. Do not attempt to open the cabinet.

Source: Phoenix Contact WP-G HF instructions

Network cable
A standard network cable can be fed through the cable entry port on the table side.
An internet connection is recommended to enable remote support. See the IT chapter for more details.
If a fixed internet connection is not available on site, you may be able to use a Wi-Fi bridge or a 5G network modem. Check with your IT department or network administrator.
We advise against using USB-to-Wi-Fi adapters, as they may not provide a stable or reliable connection because of the metal cabinet or the need for USB extension cables.
Do not interfere with or tamper with the purple cables or the internal network switch.
This network is reserved exclusively for internal components. Interference may cause damage or pose a safety risk.
Disassembly
To disassemble the EST for transport, follow the assembly instructions above in reverse.
- Ensure that the EST is disconnected from the mains before you start work. Switching it off is not sufficient. See Safety Warnings.
- Ensure that all cables between the modules being separated are properly disconnected and labelled.
- Keep all screws and fasteners in a safe place to avoid losing them during transport.