Technical Specifications
Below are the technical specifications of the EST. These specifications are based on the standard configuration of the EST, which consists of 3 modules: 1x Cable unit, 1x Cabinet and a Table.
Ambient temperature range: 15 – 40 °C Storage temperature range: -10 – 60 °C
These specifications are valid after the EST has been powered on for at least 30 minutes.
Dimensions
A standard EST Measures 2.6 x 2.0 x 0.7* meters (width x height x depth). The dimensions of the EST are shown in the figure below. If ordered using a different configuration, multiply the dimensions per module accordingly.
*frame only, total depth including plugs is approx 0.9 m.
Weight
Standard configuration weighs approximately 375 kg.
| Module | Weight (kg) |
|---|---|
| Cable unit | 125 |
| Cabinet | 200 |
| Table | 50 |
Power Rating
The EST CEE 16A 5-pin CEE power supply input is on the table side on the lowest shelf.
| Specification | Value |
|---|---|
| Inlet | CEE 16A 5-pin |
| Voltage | 400/230V |
| Frequency | 50 Hz |
| Current | 16 A |
| Power (VA) | 1000VA |
| Power (W) | 500W* |
| Internal circuit breaker | C10A |
| Max external circuit breaker | C16A |
*Typical usage 125 W.
Measurement specifications
Summarized overview of the measurements performed in the EST, their resolution and accuracy.
Some test ranges result in better performance. For more details see the sections below.
| Measurement | Resolution | Accuracy | Notes |
|---|---|---|---|
| Connection | yes/no | 40 Ohm | 1mA, 6.5V |
| Resistance of conductors | 100 μΩ | ±(0.05% of reading+ 0.02% of range) | 10mA, 6.5V |
| Resistance of PE conductors | 0.01 Ω | ±(2% of reading +2 digits) | 0.2A, 6.5V |
| Insulation resistance | 0.1 MΩ | ±10 % of reading | test voltage -0 %, +20 % |
| RCD t | 0.1 ms | ±3 ms | current within 0.1 IΔN |
| RCD I | 0.05 xIΔN | ±0.1 xIΔN | current within 0.1 IΔN |
| Voltage | 1 V | ±(2 % of reading + 2 digits) |
Functional test power source
The EST contains a three phase 400V/230V power transformer for a 100VA power source for RCD and Functional tests.
This power source is not isolated.
Instruments
The EST used multiple instruments to perform the required measurements. Specific instruments used in the EST and their specifications are described below. Not all specifications may be included, for more details see the user manuals of the instruments.
Some instruments may not be present in all EST configurations.
GW-Instek GOM804
This instruments is only used for continuity check (open/short) and resistance measurements for circuits not classified as PhasePE.
Resistance
| Resistance range | Resolution | Current | Accuracy | open circuit voltage |
|---|---|---|---|---|
| <50mΩ | 1μΩ | 1A | ±(0.1% of reading+ 0.02% of range) | 6.5V |
| <500mΩ | 10μΩ | 100mA | ±(0.05% of reading+ 0.02% of range) | 6.5V |
| <5Ω | 100μΩ | 10mA | ±(0.05% of reading+ 0.02% of range) | 6.5V |
Metrel MI3325
This instrument is used for (insulation) resistance measurements and RCD testing.
No results are stored in the instrument.
Protective Conductor Resistance (Continuity_EE)
This instrument is used for resistance measurements of circuits classified as PhasePE.
| Resistance range | Current | Accuracy | open circuit voltage |
|---|---|---|---|
| <19.99 Ω | 0.2A* | ±(2% of reading +2D) | 6 Vac |
*If resistance <8 Ohm
Insulation Resistance (Uiso_INST_EE)
Insulation resistance is measured at different test voltages, depending on the test object and the test being performed.
| Range | Resolution | Accuracy Uiso: >=500V |
|---|---|---|
| 0.15 MΩ … 19.99 MΩ | 0.01 MΩ | ±(5 % of reading + 3 digits) |
| 20.0 MΩ … 99.9 MΩ | 0.1 MΩ | ±5 % of reading |
| 100.0 MΩ … 199.9 MΩ | 1 MΩ | ±10 % of reading |
Test Voltage
| Range (V) | Resolution (V) | Accuracy |
|---|---|---|
| 0 V … 1200 V | 1 V | ±(3 % of reading + 3 digits) |
Specifications are summarized, for more details and for see the user manual of the Metrel MI3325.
Open circuit voltage: -0 %, +20 % of nominal voltage
Measuring current: min. 1 mA at Rₙ = Uₙ x 1 kΩ/V
Short circuit current: max. 3 mA
The circuit is discharged after test
In case of high moisture the measurement may be impaired. Ensure the internal heaters remain operational.
RCD testing
Residual current accuracy
| Current | Accuracy |
|---|---|
| 0.5xIΔN | -0.1·IΔ / +0 |
| IΔN, 2xIΔN, 5xIΔN | 0 / +0.1·IΔ |
Nominal residual current (A, AC):
10 mA, 15 mA, 30 mA, 100 mA, 300 mA, 500 mA, 1000 mA
Test current shape:
Sine-wave (AC), pulsed (A, F), smooth DC (B, B+)
Not all RCD tests are available for >0.3A IΔN RCDs.
RCD t - trip time
| Range | Resolution | Accuracy |
|---|---|---|
| 0.0 - 40.0 ms | 0.1 ms | ±1 ms |
| 0.0 - ... | 0.1 ms | ±3 ms |
RCD I - trip current
| Range | Resolution | Accuracy |
|---|---|---|
| 0.2 ... 1.1 xIΔN (AC-type) | 0.05 xIΔN | ±0.1 xIΔN |
| 0.2 ... 1.5 xIΔN (A-type >30mA) | 0.05 xIΔN | ±0.1 xIΔN |
| 0.2 ... 2.2 xIΔN (A-type <30mA) | 0.05 xIΔN | ±0.1 xIΔN |
| 0.2 ... 2.2 xIΔN (B-type) | 0.05 xIΔN | ±0.1 xIΔN |
Uc (contact voltage)
| Range | Resolution | Accuracy |
|---|---|---|
| 0.0–19.9 V | 0.1 V | (-0 / +15 %) ±10 digits |
| 20.0–99.9 V | 0.1 V | (-0 / +15 %) |
Voltage
In some tests the voltage is measured.
| Range | Resolution | Accuracy |
|---|---|---|
| 0.0 V … 99.9 V | 0.1 V | |
| 100 V … 550 V | 1 V | ±(2 % of reading + 2 digits) |