Technical Specifications

Below are the technical specifications of the EST. These specifications are based on the standard configuration of the EST, which consists of 3 modules: 1x Cable unit, 1x Cabinet and a Table.

Ambient temperature range: 15 – 40 °C Storage temperature range: -10 – 60 °C

These specifications are valid after the EST has been powered on for at least 30 minutes.

Dimensions​

A standard EST Measures 2.6 x 2.0 x 0.7* meters (width x height x depth). The dimensions of the EST are shown in the figure below. If ordered using a different configuration, multiply the dimensions per module accordingly.

*frame only, total depth including plugs is approx 0.9 m.

Dimensions of the EST

Weight​

Standard configuration weighs approximately 375 kg.

ModuleWeight (kg)
Cable unit125
Cabinet200
Table50

Power Rating​

The EST CEE 16A 5-pin CEE power supply input is on the table side on the lowest shelf.

SpecificationValue
InletCEE 16A 5-pin
Voltage400/230V
Frequency50 Hz
Current16 A
Power (VA)1000VA
Power (W)500W*
Internal circuit breakerC10A
Max external circuit breakerC16A

*Typical usage 125 W.

Measurement specifications​

Summarized overview of the measurements performed in the EST, their resolution and accuracy.
Some test ranges result in better performance. For more details see the sections below.

MeasurementResolutionAccuracyNotes
Connectionyes/no40 Ohm1mA, 6.5V
Resistance of conductors100 μΩ±(0.05% of reading+ 0.02% of range)10mA, 6.5V
Resistance of PE conductors0.01 Ω±(2% of reading +2 digits)0.2A, 6.5V
Insulation resistance0.1 MΩ±10 % of readingtest voltage -0 %, +20 %
RCD t0.1 ms±3 mscurrent within 0.1 IΔN
RCD I0.05 xIΔN±0.1 xIΔNcurrent within 0.1 IΔN
Voltage1 V±(2 % of reading + 2 digits)

Functional test power source​

The EST contains a three phase 400V/230V power transformer for a 100VA power source for RCD and Functional tests.

danger

This power source is not isolated.


Instruments​

The EST used multiple instruments to perform the required measurements. Specific instruments used in the EST and their specifications are described below. Not all specifications may be included, for more details see the user manuals of the instruments.

Some instruments may not be present in all EST configurations.

GW-Instek GOM804​

This instruments is only used for continuity check (open/short) and resistance measurements for circuits not classified as PhasePE.

Resistance​

Resistance rangeResolutionCurrentAccuracyopen circuit voltage
<50mΩ1μΩ1A±(0.1% of reading+ 0.02% of range)6.5V
<500mΩ10μΩ100mA±(0.05% of reading+ 0.02% of range)6.5V
<5Ω100μΩ10mA±(0.05% of reading+ 0.02% of range)6.5V

Metrel MI3325​

This instrument is used for (insulation) resistance measurements and RCD testing.

note

No results are stored in the instrument.

Protective Conductor Resistance (Continuity_EE)​

This instrument is used for resistance measurements of circuits classified as PhasePE.

Resistance rangeCurrentAccuracyopen circuit voltage
<19.99 Ω0.2A*±(2% of reading +2D)6 Vac

*If resistance <8 Ohm

Insulation Resistance (Uiso_INST_EE)​

Insulation resistance is measured at different test voltages, depending on the test object and the test being performed.

RangeResolutionAccuracy Uiso: >=500V
0.15 MΩ … 19.99 MΩ0.01 MΩ±(5 % of reading + 3 digits)
20.0 MΩ … 99.9 MΩ0.1 MΩ±5 % of reading
100.0 MΩ … 199.9 MΩ1 MΩ±10 % of reading
Test Voltage​
Range (V)Resolution (V)Accuracy
0 V … 1200 V1 V±(3 % of reading + 3 digits)
note

Specifications are summarized, for more details and for see the user manual of the Metrel MI3325.

Open circuit voltage: -0 %, +20 % of nominal voltage
Measuring current: min. 1 mA at Rₙ = Uₙ x 1 kΩ/V
Short circuit current: max. 3 mA
The circuit is discharged after test

note

In case of high moisture the measurement may be impaired. Ensure the internal heaters remain operational.


RCD testing​

Residual current accuracy

CurrentAccuracy
0.5xIΔN-0.1·IΔ / +0
IΔN, 2xIΔN, 5xIΔN0 / +0.1·IΔ

Nominal residual current (A, AC):
10 mA, 15 mA, 30 mA, 100 mA, 300 mA, 500 mA, 1000 mA

Test current shape:
Sine-wave (AC), pulsed (A, F), smooth DC (B, B+)

Not all RCD tests are available for >0.3A IΔN RCDs.

RCD t - trip time​

RangeResolutionAccuracy
0.0 - 40.0 ms0.1 ms±1 ms
0.0 - ...0.1 ms±3 ms

RCD I - trip current​

RangeResolutionAccuracy
0.2 ... 1.1 xIΔN (AC-type)0.05 xIΔN±0.1 xIΔN
0.2 ... 1.5 xIΔN (A-type >30mA)0.05 xIΔN±0.1 xIΔN
0.2 ... 2.2 xIΔN (A-type <30mA)0.05 xIΔN±0.1 xIΔN
0.2 ... 2.2 xIΔN (B-type)0.05 xIΔN±0.1 xIΔN

Uc (contact voltage)​

RangeResolutionAccuracy
0.0–19.9 V0.1 V(-0 / +15 %) ±10 digits
20.0–99.9 V0.1 V(-0 / +15 %)

Voltage​

In some tests the voltage is measured.

RangeResolutionAccuracy
0.0 V … 99.9 V0.1 V
100 V … 550 V1 V±(2 % of reading + 2 digits)