Safety Warnings
The EST is an automated electrical safety tester.
During operation voltages up to 1000 Vdc / 400 Vac are applied on the test connections.
Never touch exposed parts of test objects during a test. Observe the safety light.
Do not use the EST if any part of it is damaged.
Do not connect any test connections of the EST or test object connected to the EST to mains or any external power supply.
Use personal protection when working with live equipment. (EN 60903-400V)
Do not modify or extend the functionality of the EST unless approved by EMRI or the user manual.
Special condition apply to Connection Overrides, refer to Connection Overrides.
The EST is equipped with heaters in all cabinets to avoid condensation and oxidation.
The heaters remain live even when the power switch if off.
Do not modify or extend the internal ethernet network (purple cables)
Do not tamper with software and or databases, doing so may cause malfunctions or safety hazards.
Do not expose the software to the internet.
Do not climb or stand on the EST
Do not use the EST cabinets for storage
Toolbox Meeting
It is recommended to hold a toolbox meeting with all personnel involved in the use of the EST.
The purpose of this meeting is to ensure that everyone understands the hazards associated with the EST and the equipment being tested, as well as the safety procedures that must be followed.
Electrical equipment being tested may have defects and may be dangerous.
Hazards of the EST
Although efforts are made to ensure the EST is safe, some tests require high voltages.
- Do not connect equipment under test to any other power source than the EST.
- The EST will perform an insulation test with up to 1000Vdc, do not touch the equipment under test during this test unless explicitly instructed to do so.
- The EST will perform a functional RCD test with up to 400Vac, do not touch the equipment under test during this test unless explicitly instructed to do so.
- Do not create dangerous custom adapters, for example with male-to-male connectors.
Observe the safety light and follow the instructions of the EST.
Hazards of equipment under test
Various types of equipment can be tested with the EST. Overall you should not connect equipment to the EST that is in bad shape.
Dangers of equipment under test may vary depending on the type of equipment and are the responsibility of the user to assess.
- Do not connect equipment that is damaged or has exposed live parts.
- Do not connect equipment that is out of scope for the EST, only test distribution boxes and cables.
- Do not repair or modify equipment under test while it is connected to the EST.