Power
Power Inlet
The EST is powered by a single CEE power inlet located on the table side of the EST.
See Technical Specifications for details.
Main isolator/breaker
The main isolator/breaker is located inside the cabinet on the bottom left.
In off position, all power is cut off, including the internal heating.
When plugged in the internal heaters inside always work regardless of the power switch state to keep the EST dry from condensation.
Power switch
The power switch controls the measurement system and computer.
On
Turn on the EST with the power switch on the table side. The computer will automatically start on power-on.
Windows will take 3 minutes to be completely ready after login.
Off
To turn off the EST, first shutdown the computer by clicking the Windows Start button and selecting "Shut down".
To avoid data loss, wait for the computer to fully shut down before turning off the main power switch.